The JHM710A Infusion Pump is a medical-grade precision device designed to deliver safe, accurate fluid and medication infusion for diverse clinical scenarios. Adopting a classic vertical structure and dual CPU design, it ensures reliable performance while simplifying operation—equipped with a 3.2-inch high-brightness LCD screen for clear data visualization and a numeric keypad for quick parameter input, reducing 医护人员 (medical staff) workload and operational errors.
Item Specification
Infusion Mode Volume Mode, Time Mode, Drug Mode
Rate Unit ml/h, volume/time, d/min
IV Sets Standard single use IV sets (10, 15, 20, 60 d/mL) with tube diameter 3.4 ~ 4.5 mm
Flow Rate 1.0 ml/h ~ 1200 ml/h; 1.0 d/min ~ 400 d/min
Increment 1.0 – 99.9 ml/h: 0.1 ml/h step; 100 – 1200 ml/h: 1 ml/h step; 1 – 400 d/min: 1 d/min step
Deviation Flow Rate: < 5% (after calibration)
VTBI (Volume to Be Infused) 0.1 ~ 9999.9 ml
Total Volume 0.0 ~ 9999.9 ml
Time Setting 1 min ~ 99 h 00 min
Purge Rate 600 ml/h, display purge volume
KVO (Keep Vein Open) 1.0 ml/h ~ 5.0 ml/h, adjustable
Keypad Numeric keypad
BOLUS Function Available, flow rate adjustable
Night Mode Screen dims automatically after 5 mins of standby
Occlusion Pressure Detector Upstream and downstream pressure detection; 40 KPa ~ 130 KPa; 10 levels adjustable with dynamic pressure value display
Air Bubble Detector Ultrasonic detection; ≥ 50 μl; Single bubble volume: 50 ~ 300 μl; 10 levels adjustable
Alarm Volume 10 levels adjustable
Motor Anti-reverse Function Equipped to prevent upstream fluid backflow
IV Set Management Open system; compatible with different IV set brands after calibration
Self-Testing Pre-startup self-test and in-infusion self-test
Audio Visual Alarm Messages Air bubble, upstream occlusion, downstream occlusion, door open, infusion finish, standby time expired, battery low, battery empty, battery disengaged, AC Power disengaged, motor error, pump error, pressure sensor error, parameter error, system error
Screen 3.2-inch high-brightness LCD
Power Supply AC Power: 100 V ~ 240 V, 50 Hz / 60 Hz, ≤ 25 VA
Battery Rechargeable Lithium battery; DC 11.1 V, 3000 mAh; Operating time: > 10 h @ 25 ml/h; Displays remaining battery
Operating Conditions Temperature: +5°C ~ +40°C; Humidity: 20% ~ 90%; Air pressure: 70 ~ 106 KPa
Transport & Storage Conditions Temperature: -20°C ~ +55°C; Humidity: 10% ~ 95%; Air pressure: 50 ~ 106 KPa
Device Classification Type CF, Class II, IPX3
Dimension & Weight Dimension: 145 mm × 150 mm × 200 mm; Weight: 1.7 kg